1 Introduction

Data quality assessment is an integral part of preparatory data analysis to ensure sound biological information retrieval.

We present here the MsQuality package, which provides functionality to calculate quality metrics for mass spectrometry-derived, spectral data at the per-sample level. MsQuality relies on the mzQC framework of quality metrics defined by the Human Proteome Organization-Proteomics Standards Intitiative (HUPO-PSI). These metrics quantify the quality of spectral raw files using a controlled vocabulary. The package is especially addressed towards users that acquire mass spectrometry data on a large scale (e.g. data sets from clinical settings consisting of several thousands of samples): while it is easier to control for high-quality data acquisition in small-scale experiments, typically run in one or few batches, clinical data sets are often acquired over longer time frames and are prone to higher technical variation that is often unnoticed. MsQuality tries to address this problem by calculating metrics that can be stored along the spectral data sets (raw files or feature-extracted data sets). MsQuality, thus, facilitates the tracking of shifts in data quality and quantifies the quality using multiple metrics. It should be thus easier to identify samples that are of low quality (high-number of missing values, termination of chromatographic runs, low instrument sensitivity, etc.).

We would like to note here that these metrics only give an indication of data quality, and, before removing indicated low-quality samples from the analysis more advanced analytics, e.g. using the implemented functionality and visualizations in the MatrixQCvis package, should be scrutinized. Also, data quality should always be regarded in the context of the sample type and experimental settings, i.e. quality metrics should always be compared with regard to the sample type, experimental setup, instrumentation, etc..

The MsQuality package allows to calculate low-level quality metrics that require minimum information on mass spectrometry data: retention time, m/z values, and associated intensities. The list included in the mzQC framework is excessive, also including metrics that rely on more high-level information, that might not be readily accessible from .raw or .mzML files, e.g. pump pressure mean, or rely on alignment results, e.g. retention time mean shift, signal-to-noise ratio, precursor errors (ppm).

The MsQuality package is built upon the Spectra and the MsExperiment package. Metrics will be calculated based on the information stored in a Spectra object and the respective dataOrigin entries are used to distinguish between the mass spectral data of multiple samples. The MsExperiment serves as a container to store the mass spectral data of multiple samples. MsQuality enables the user to calculate quality metrics both on Spectra and MsExperiment objects.

MsQuality can be used for any type of experiment that can be represented as a Spectra or MsExperiment object. This includes simple LC-MS data, DIA or DDA-based data, ion mobility data or MS data in general. The tool can thus be used for any type of targeted or untargeted metabolomics or proteomics workflow. Also, we are not limited to data files in mzML format, but, through Spectra and related MsBackend packages, data can be imported from a large variety of formats, including some raw vendor formats.

In this vignette, we will (i) create some exemplary Spectra and MsExperiment objects, (ii) calculate the quality metrics on these data sets, and (iii) visualize some of the metrics.

1.1 Alternative software for data quality assessment

Other R packages are available in Bioconductor that are able to assess the quality of mass spectrometry data:

artMS uses MaxQuant output and enables to calculate several QC metrics, e.g. correlation matrix for technical replicates, calculation of total sum of intensities in biological replicates, total peptide counts in biological replicates, charge state distribution of PSMs identified in each biological replicates, or MS1 scan counts in each biological replicate.

MSstatsQC and the visualization tool MSstatsQCgui require csv files in long format from spectral processing tools such as Skyline and Panorama autoQC or MSnbase objects. MSstatsQC enables to generate individual, moving range, cumulative sum for mean, and/or cumulative sum for variability control charts for each metric. Metrics can be any kind of user-defined metric stored in the data columns for a given peptide, e.g. retention time and peak area.

MQmetrics provides a pipeline to analyze the quality of proteomics data sets from MaxQuant files and focuses on proteomics-/MaxQuant-specific metrics, e.g. proteins identified, peptides identified, or proteins versus peptide/protein ratio.

MatrixQCvis provides an interactive shiny-based interface to assess data quality at various processing steps (normalization, transformation, batch correction, and imputation) of rectangular matrices. The package includes several diagnostic plots and metrics such as barplots of intensity distributions, plots to visualize drifts, MA plots and Hoeffding’s D value calculation, and dimension reduction plots and provides specific tools to analyze data sets containing missing values as commonly observed in mass spectrometry.

proBatch enables to assess batch effects in (prote)omics data sets and corrects these batch effects in subsequent steps. Several tools to visualize data quality are included in the proBatch packages, such as barplots of intensity distributions, cluster and heatmap analysis tools, and PCA dimension reduction plots. Additionally, proBatch enables to assess diagnostics at the feature level, e.g. peptides or spike-ins.

2 Installation

To install this package, start R and enter:

if (!requireNamespace("BiocManager", quietly = TRUE))
    install.packages("BiocManager")

if (!requireNamespace("remotes", quietly = TRUE))
    install.packages("remotes")

## to install from Bioconductor
BiocManager::install("MsQuality")

## to install from GitHub
BiocManager::install("tnaake/MsQuality")

This will install this package and all eventually missing dependencies.

Questions and bugs

MsQuality is currently under active development. If you discover any bugs, typos or develop ideas of improving MsQuality feel free to raise an issue via GitHub or send a mail to the developer.

3 Create Spectra and MsExperiment objects

Load the Spectra package.

library("Spectra")
library("MsExperiment")
## Loading required package: ProtGenerics
## 
## Attaching package: 'ProtGenerics'
## The following object is masked from 'package:stats':
## 
##     smooth
library("MsQuality")

3.1 Create Spectra and MsExperiment objects from mzML files

There are several options available to create a Spectra object. One way, as outlined in the vignette of the Spectra package is by specifying the location of mass spectrometry raw files in mzML, mzXML or CDF format and using the MsBackendMzR backend. Here we load the example files from the sciex data set of the msdata package and create a Spectra object from the two provided mzML files. The example is taken from the Spectra vignette.

## this example is taken from the Spectra vignette
fls <- dir(system.file("sciex", package = "msdata"), full.names = TRUE)
sps_sciex <- Spectra(fls, backend = MsBackendMzR())

The data set consists of a single sample measured in two different injections to the same LC-MS setup. An empty instance of an MsExperiment object is created and populated with information on the samples by assigning data on the samples (sampleData), information on the mzML files (MsExperimentFiles) and spectral information (spectra). In a last step, using linkSampleData, the relationships between the samples and the spectral information are defined.

## this example is taken from the Spectra vignette
lmse <- MsExperiment()
sd <- DataFrame(sample_id = c("QC1", "QC2"),
    sample_name = c("QC Pool", "QC Pool"),
    injection_idx = c(1, 3))
sampleData(lmse) <- sd
## add mzML files to the experiment
experimentFiles(lmse) <- MsExperimentFiles(mzML_files = fls)
## add the Spectra object to the experiment
spectra(lmse) <- sps_sciex
## use linkSampleData to establish and define relationships between sample 
## annotations and MS data
lmse <- linkSampleData(lmse, with = "experimentFiles.mzML_file",
    sampleIndex = c(1, 2), withIndex = c(1, 2))

3.2 Create Spectra and MsExperiment objects from (feature-extracted) intensity tables

Another common approach is the creation of Spectra objects from a DataFrames using the MsBackendDataFrame backend.

We will use here the data set of Lee et al. (2019), that contains metabolite level information measured by reverse phase liquid chromatography (RPLC) coupled to mass spectrometry and hydrophilic interaction liquid chromatography (HILIC) coupled to mass spectrometry (derived from the file STables - rev1.xlsx in the Supplementary Information).

In a separate step (see documentation for Lee2019_meta_vals and Lee2019), we have created a list containing Spectra objects for each samples (objects sps_l_rplc and sps_l_hilic) and MsExperiment objects containing the data of all samples (objects msexp_rplc and msexp_hilic). We will load here these objects:

data("Lee2019", package = "MsQuality")

The final data set contains 541 paired samples (i.e. 541 samples derived from RPLC and 541 samples derived from HILIC).

We will combine the sps_rplc and sps_hilic objects in the following and calculate on this combined document the metrics.

sps_comb <- c(sps_rplc, sps_hilic)

The most important function to assess the data quality and to calculate the metrics is the calculateMetrics function. The function takes a Spectra or MsExperiment object as input, a character vector of metrics to be calculated, and, optionally a list of parameters passed to the quality metrics functions.

4 Calculating the quality metrics on Spectra and MsExperiment objects

Currently, the following metrics are included:

qualityMetrics(sps_comb)
##  [1] "chromatographyDuration"             "ticQuartersRtFraction"             
##  [3] "rtOverMsQuarters"                   "ticQuartileToQuartileLogRatio"     
##  [5] "numberSpectra"                      "numberEmptyScans"                  
##  [7] "medianPrecursorMz"                  "rtIqr"                             
##  [9] "rtIqrRate"                          "areaUnderTic"                      
## [11] "areaUnderTicRtQuantiles"            "extentIdentifiedPrecursorIntensity"
## [13] "medianTicRtIqr"                     "medianTicOfRtRange"                
## [15] "mzAcquisitionRange"                 "rtAcquisitionRange"                
## [17] "precursorIntensityRange"            "precursorIntensityQuartiles"       
## [19] "precursorIntensityMean"             "precursorIntensitySd"              
## [21] "msSignal10xChange"                  "ratioCharge1over2"                 
## [23] "ratioCharge3over2"                  "ratioCharge4over2"                 
## [25] "meanCharge"                         "medianCharge"

4.1 List of included metrics

The following list gives a brief explanation on the included metrics. Further information may be found at the HUPO-PSI mzQC project page or in the respective help file for the quality metric (accessible by e.g. entering ?chromatographyDuration to the R console). We also give here explanation on how the metric is calculated in MsQuality. Currently, all quality metrics can be calculated for both Spectra and MsExperiment objects.

  • chromatographyDuration, chromatography duration (MS:4000053), “The retention time duration of the chromatography in seconds.” [PSI:MS]; Longer duration may indicate a better chromatographic separation of compounds which depends, however, also on the sampling/scan rate of the MS instrument.

    The metric is calculated as follows:

    1. the retention time associated to the Spectra object is obtained,
    2. the maximum and the minimum of the retention time is obtained,
    3. the difference between the maximum and the minimum is calculated and returned.
  • ticQuartersRtFraction, TIC quarters RT fraction (MS:4000054), “The interval when the respective quarter of the TIC accumulates divided by retention time duration.” [PSI:MS]; The metric informs about the dynamic range of the acquisition along the chromatographic separation. The metric provides information on the sample (compound) flow along the chromatographic run, potentially revealing poor chromatographic performance, such as the absence of a signal for a significant portion of the run.

    The metric is calculated as follows:

    1. the Spectra object is ordered according to the retention time,
    2. the cumulative sum of the ion count is calculated (TIC),
    3. the quantiles are calculated according to the probs argument, e.g. when probs is set to c(0, 0.25, 0.5, 0.75, 1) the 0%, 25%, 50%, 75%, and 100% quantile is calculated,
    4. the retention time/relative retention time (retention time divided by the total run time taking into account the minimum retention time) is calculated,
    5. the (relative) duration of the LC run after which the cumulative TIC exceeds (for the first time) the respective quantile of the cumulative TIC is calculated and returned.
  • rtOverMsQuarters, MS1 quarter RT fraction (MS:4000055), “The interval used for acquisition of the first, second, third, and fourth quarter of all MS1 events divided by retention time duration.” [PSI:MS], msLevel = 1L; The metric informs about the dynamic range of the acquisition along the chromatographic separation. For MS1 scans, the values are expected to be in a similar range across samples of the same type.

    The metric is calculated as follows:

    1. the retention time duration of the whole Spectra object is determined (taking into account all the MS levels),
    2. the Spectra object is filtered according to the MS level and subsequently ordered according to the retention time,
    3. the MS events are split into four (approximately) equal parts,
    4. the relative retention time is calculated (using the retention time duration from (1) and taking into account the minimum retention time),
    5. the relative retention time values associated to the MS event parts are returned.
  • rtOverMsQuarters, MS2 quarter RT fraction (MS:4000056), “The interval used for acquisition of the first, second, third, and fourth quarter of all MS2 events divided by retention time duration.” [PSI:MS], msLevel = 2L; The metric informs about the dynamic range of the acquisition along the chromatographic separation. For MS2 scans, the comparability of the values depends on the acquisition mode and settings to select ions for fragmentation.

    The metric is calculated as follows:

    1. the retention time duration of the whole Spectra object is determined (taking into account all the MS levels),
    2. the Spectra object is filtered according to the MS level and subsequently ordered according to the retention time,
    3. the MS events are split into four (approximately) equal parts,
    4. the relative retention time is calculated (using the retention time duration from (1) and taking into account the minimum retention time),
    5. the relative retention time values associated to the MS event parts are returned.
  • ticQuartileToQuartileLogRatio, MS1 TIC-change quartile ratios (MS:4000057), ““The log ratios of successive TIC-change quartiles. The TIC changes are the list of MS1 total ion current (TIC) value changes from one to the next scan, produced when each MS1 TIC is subtracted from the preceding MS1 TIC. The metric’s value triplet represents the log ratio of the TIC-change Q2 to Q1, Q3 to Q2, TIC-change-max to Q3” [PSI:MS], mode = "TIC_change", relativeTo = "previous", msLevel = 1L; The metric informs about the dynamic range of the acquisition along the chromatographic separation.This metric evaluates the stability (similarity) of MS1 TIC values from scan to scan along the LC run. High log ratios representing very large intensity differences between pairs of scans might be due to electrospray instability or presence of a chemical contaminant.

    The metric is calculated as follows:

    1. the TIC (ionCount) of the Spectra object is calculated per scan event (with spectra ordered by retention time),
    2. the differences between TIC values are calculated between subsequent scan events,
    3. the ratios between the 25%, 50%, 75%, and 100% quantile to the 25% quantile of the values of (2) are calculated,
    4. the log values of the ratios are returned.
  • ticQuartileToQuartileLogRatio, MS1 TIC quartile ratios (MS:4000058), “The log ratios of successive TIC quartiles. The metric’s value triplet represents the log ratios of TIC-Q2 to TIC-Q1, TIC-Q3 to TIC-Q2, TIC-max to TIC-Q3.” [PSI:MS], mode = "TIC", relativeTo = "previous", msLevel = 1L; The metric informs about the dynamic range of the acquisition along the chromatographic separation. The ratios provide information on the distribution of the TIC values for one LC-MS run. Within an experiment, with the same LC setup, values should be comparable between samples.

    The metric is calculated as follows:

    1. the TIC (ionCount) of the Spectra object is calculated per scan event (with spectra ordered by retention time),
    2. the TIC values between subsequent scan events are taken as they are,
    3. the ratios between the 25%, 50%, 75%, and 100% quantile to the 25% quantile of the values of (2) are calculated.
    4. The log values of the ratios are returned.
  • numberSpectra, number of MS1 spectra MS:4000059), “The number of MS1 events in the run.” [PSI:MS], msLevel = 1L; An unusual low number may indicate incomplete sampling/scan rate of the MS instrument, low sample volume and/or failed injection of a sample.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the number of the spectra are obtained (length of Spectra) and returned.
  • numberSpectra, number of MS2 spectra (MS:4000060), “The number of MS2 events in the run.” [PSI:MS], msLevel = 2L; An unusual low number may indicate incomplete sampling/scan rate of the MS instrument, low sample volume and/or failed injection of a sample.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the number of the spectra are obtained (length of Spectra) and returned.
  • mzAcquisitionRange, m/z acquisition range (MS:4000069), “Upper and lower limit of m/z precursor values at which MSn spectra are recorded.” [PSI:MS]; The metric informs about the dynamic range of the acquisition. Based on the used MS instrument configuration, the values should be similar. Variations between measurements may arise when employing acquisition in DDA mode.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the m/z values of the peaks within the Spectra object are obtained,
    3. the minimum and maximum m/z values are obtained and returned.
  • rtAcquisitionRange, retention time acquisition range (MS:4000070), “Upper and lower limit of retention time at which spectra are recorded.” [PSI:MS]; An unusual low range may indicate incomplete sampling and/or a premature or failed LC run.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the retention time values of the features within the Spectra object are obtained,
    3. the minimum and maximum retention time values are obtained and returned.
  • msSignal10xChange, MS1 signal jump (10x) count (MS:4000097), “The number of times where MS1 TIC increased more than 10-fold between adjacent MS1 scans. An unusual high count of signal jumps or falls can indicate ESI stability issues.” [PSI:MS], change = "jump", msLevel = 1L; An unusual high count of signal jumps or falls may indicate ESI stability issues.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the intensity values of the features are obtained via the ion count,
    4. the signal jumps/declines of the intensity values with the two subsequent intensity values is calculated,
    5. the signal jumps by a factor of ten or more are counted and returned.
  • msSignal10xChange, MS1 signal fall (10x) count (MS:4000098), “The number of times where MS1 TIC decreased more than 10-fold between adjacent MS1 scans. An unusual high count of signal jumps or falls can indicate ESI stability issues.” [PSI:MS], change = "fall", msLevel = 1L; An unusual high count of signal jumps or falls may indicate ESI stability issues.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the intensity values of the features are obtained via the ion count,
    4. the signal jumps/declines of the intensity values with the two subsequent intensity values is calculated,
    5. the signal declines by a factor of ten or more are counted and returned.
  • numberEmptyScans, number of empty MS1 scans (MS:4000099), “Number of MS1 scans where the scans’ peaks intensity sums to 0 (i.e. no peaks or only 0-intensity peaks).” [PSI:MS], msLevel = 1L; An unusual high number may indicate incomplete sampling/scan rate of the MS instrument, low sample volume and/or failed injection of a sample.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensities per entry are obtained,
    3. the number of intensity entries that are NULL, NA, or that have a sum of 0 are obtained and returned.
  • numberEmptyScans, number of empty MS2 scans (MS:4000100), “Number of MS2 scans where the scans’ peaks intensity sums to 0 (i.e. no peaks or only 0-intensity peaks).” [PSI:MS], msLevel = 2L; An unusual high number may indicate incomplete sampling/scan rate of the MS instrument, low sample volume and/or failed injection of a sample.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensities per entry are obtained,
    3. the number of intensity entries that are NULL, NA, or that have a sum of 0 are obtained and returned.
  • numberEmptyScans, number of empty MS3 scans (MS:4000101), “Number of MS3 scans where the scans’ peaks intensity sums to 0 (i.e. no peaks or only 0-intensity peaks).” [PSI:MS], msLevel = 3L; An unusual high number may indicate incomplete sampling/scan rate of the MS instrument, low sample volume and/or failed injection of a sample.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensities per entry are obtained,
    3. the number of intensity entries that are NULL, NA, or that have a sum of 0 are obtained and returned.
  • precursorIntensityQuartiles, MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000116), “From the distribution of MS2 precursor intensities, the quartiles Q1, Q2, Q3.” [PSI:MS], identificationLevel = "all"; The intensity distribution of the precursors informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the 25%, 50%, and 75% quantile of the precursor intensity values are obtained (NA values are removed) and returned.
  • precursorIntensityMean, MS2 precursor intensity distribution mean (MS:4000117), “From the distribution of MS2 precursor intensities, the mean.” [PSI:MS], identificationLevel = "all"; The intensity distribution of the precursors informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the mean of the precursor intensity values is obtained (NA values are removed) and returned.
  • precursorIntensitySd, MS2 precursor intensity distribution sigma (MS:4000118), “From the distribution of MS2 precursor intensities, the sigma value.” [PSI:MS], identificationLevel = "all"; The intensity distribution of the precursors informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the standard deviation of precursor intensity values is obtained (NA values are removed) and returned.
  • medianPrecursorMz, MS2 precursor median m/z of identified quantification data points (MS:4000152), “Median m/z value for MS2 precursors of all quantification data points after user-defined acceptance criteria are applied. These data points may be for example XIC profiles, isotopic pattern areas, or reporter ions (see MS:1001805). The used type should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified", msLevel = 1L; The m/z distribution informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor m/z values are obtained,
    3. the median value is returned (NAs are removed).
  • rtIqr, interquartile RT period for identified quantification data points (MS:4000153), “The interquartile retention time period, in seconds, for all quantification data points after user-defined acceptance criteria are applied over the complete run. These data points may be for example XIC profiles, isotopic pattern areas, or reporter ions (see MS:1001805). The used type should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Longer duration may indicate a better chromatographic separation of compounds which depends, however, also on the sampling/scan rate of the MS instrument.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the retention time values are obtained,
    3. the interquartile range is obtained from the values and returned (NA values are removed).
  • rtIqrRate, rate of the interquartile RT period for identified quantification data points (MS:4000154), “The rate of identified quantification data points for the interquartile retention time period, in identified quantification data points per second. These data points may be for example XIC profiles, isotopic pattern areas, or reporter ions (see MS:1001805). The used type should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Higher rates may indicate a more efficient sampling and identification.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the retention time values are obtained,
    3. the 25% and 75% quantiles are obtained from the retention time values (NA values are removed),
    4. the number of eluted features between this 25% and 75% quantile is calculated,
    5. the number of features is divided by the interquartile range of the retention time and returned.
  • areaUnderTic, area under TIC (MS:4000155), “The area under the total ion chromatogram.” [PSI:MS]; The metric informs about the dynamic range of the acquisition. Differences between samples of an experiment may indicate differences in the dynamic range and/or in the sample content.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the sum of the ion counts are obtained and returned.
  • areaUnderTicRtQuantiles, area under TIC RT quantiles (MS:4000156), “The area under the total ion chromatogram of the retention time quantiles. Number of quantiles are given by the n-tuple.” [PSI:MS]; The metric informs about the dynamic range of the acquisition. Differences between samples of an experiment may indicate differences in the dynamic range and/or in the sample content. The metric informs about the dynamic range of the acquisition along the chromatographic separation. Differences between samples of an experiment may indicate differences in chromatographic performance, differences in the dynamic range and/or in the sample content.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the Spectra object is ordered according to the retention time,
    3. the 0%, 25%, 50%, 75%, and 100% quantiles of the retention time values are obtained,
    4. the ion count of the intervals between the 0%/25%, 25%/50%, 50%/75%, and 75%/100% are obtained,
    5. the ion counts of the intervals are summed (TIC) and the values returned.
  • extentIdentifiedPrecursorIntensity, extent of identified MS2 precursor intensity (MS:4000157), “Ratio of 95th over 5th percentile of MS2 precursor intensity for all quantification data points after user-defined acceptance criteria are applied. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensities of the precursor ions are obtained,
    3. the 5% and 95% quantile of these intensities are obtained (1NA1 values are removed),
    4. the ratio between the 95% and the 5% intensity quantile is calculated and returned.
  • medianTicRtIqr, median of TIC values in the RT range in which the middle half of quantification data points are identified (MS:4000158), “Median of TIC values in the RT range in which half of quantification data points are identified (RT values of Q1 to Q3 of identifications). These data points may be for example XIC profiles, isotopic pattern areas, or reporter ions (see MS:1001805). The used type should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition along the chromatographic separation.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the Spectra object is ordered according to the retention time,
    3. the features between the 1st and 3rd quartile are obtained (half of the features that are present in the Spectra object),
    4. the ion count of the features within the 1st and 3rd quartile is obtained,
    5. the median value of the ion count is calculated (NA values are removed) and the median value is returned.
  • medianTicOfRtRange, median of TIC values in the shortest RT range in which half of the quantification data points are identified (MS:4000159), “Median of TIC values in the shortest RT range in which half of the quantification data points are identified. These data points may be for example XIC profiles, isotopic pattern areas, or reporter ions (see MS:1001805). The used type should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition along the chromatographic separation.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the Spectra object is ordered according to the retention time,
    3. the number of features in the Spectra object is obtained and the number for half of the features is calculated,
    4. iterate through the features (always by taking the neighbouring half of features) and calculate the retention time range of the set of features,
    5. retrieve the set of features with the minimum retention time range,
    6. calculate from the set of (5) the median TIC (NA values are removed) and return it.
  • precursorIntensityRange, MS2 precursor intensity range (MS:4000160), “Minimum and maximum MS2 precursor intensity recorded.” [PSI:MS]; The metric informs about the dynamic range of the acquisition.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the minimum and maximum precursor intensity values are obtained and returned.
  • precursorIntensityQuartiles, identified MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000161), “From the distribution of identified MS2 precursor intensities, the quartiles Q1, Q2, Q3. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the 25%, 50%, and 75% quantile of the precursor intensity values are obtained (NA values are removed) and returned.
  • precursorIntensityQuartiles, unidentified MS2 precursor intensity distribution Q1, Q2, Q3 (MS:4000162), “From the distribution of unidentified MS2 precursor intensities, the quartiles Q1, Q2, Q3. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "unidentified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the 25%, 50%, and 75% quantile of the precursor intensity values are obtained (NA values are removed) and returned.
  • precursorIntensityMean, identified MS2 precursor intensity distribution mean (MS:4000163), “From the distribution of identified MS2 precursor intensities, the mean. The intensity distribution of the identified precursors informs about the dynamic range of the acquisition in relation to identifiability. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the mean of the precursor intensity values is obtained (NA values are removed) and returned.
  • precursorIntensityMean, unidentified MS2 precursor intensity distribution mean (MS:4000164), “From the distribution of unidentified MS2 precursor intensities, the mean. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "unidentified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the mean of the precursor intensity values is obtained (NA values are removed) and returned.
  • precursorIntensitySd, identified MS2 precursor intensity distribution sigma (MS:4000165), “From the distribution of identified MS2 precursor intensities, the sigma value. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the standard deviation of precursor intensity values is obtained (NA values are removed) and returned.
  • precursorIntensitySD, unidentified MS2 precursor intensity distribution sigma (MS:4000166), “From the distribution of unidentified MS2 precursor intensities, the sigma value. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "unidentified"; The metric informs about the dynamic range of the acquisition in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the intensity of the precursor ions within the Spectra object are obtained,
    3. the standard deviation of precursor intensity values is obtained (NA values are removed) and returned.
  • ratioCharge1over2, ratio of 1+ over 2+ of all MS2 known precursor charges (MS:4000167), “The ratio of 1+ over 2+ MS2 precursor charge count of all spectra.” [PSI:MS], identificationLevel = "all"; High ratios of 1+/2+ MS2 precursor charge count may indicate inefficient ionization.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 1+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • ratioCharge1over2, ratio of 1+ over 2+ of identified MS2 known precursor charges (MS:4000168), ““The ratio of 1+ over 2+ MS2 precursor charge count of identified spectra. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; High ratios of 1+/2+ MS2 precursor charge count may indicate inefficient ionization in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 1+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • ratioCharge3over2, ratio of 3+ over 2+ of all MS2 known precursor charges (MS:4000169), “The ratio of 3+ over 2+ MS2 precursor charge count of all spectra.” [PSI:MS], identificationLevel = "all"; Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.  preference for longer peptides.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 3+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • ratioCharge3over2, ratio of 3+ over 2+ of identified MS2 known precursor charges (MS:4000170), “The ratio of 3+ over 2+ MS2 precursor charge count of identified spectra. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.  preference for longer peptides in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 3+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • ratioCharge4over2, ratio of 4+ over 2+ of all MS2 known precursor charges (MS:4000171), “The ratio of 4+ over 2+ MS2 precursor charge count of all spectra.” [PSI:MS], identificationLevel = "all"; Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.  preference for longer peptides.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 4+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • ratioCharge4over2, ratio of 4+ over 2+ of identified MS2 known precursor charges (MS:4000172), “The ratio of 4+ over 2+ MS2 precursor charge count of identified spectra. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Higher ratios of 3+/2+ MS2 precursor charge count may indicate e.g.  preference for longer peptides in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the number of precursors with charge 4+ is divided by the number of precursors with charge 2+ and the ratio is returned.
  • meanCharge, mean MS2 precursor charge in all spectra (MS:4000173), “Mean MS2 precursor charge in all spectra” [PSI:MS], identificationLevel = "all"; Higher charges may indicate inefficient ionization or e.g. preference for longer peptides.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the mean of the precursor charge values is calculated and returned.
  • meanCharge, mean MS2 precursor charge in identified spectra (MS:4000174), “Mean MS2 precursor charge in identified spectra. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Higher charges may indicate inefficient ionization or e.g. preference for longer peptides in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the mean of the precursor charge values is calculated and returned.
  • medianCharge, median MS2 precursor charge in all spectra (MS:4000175), “Median MS2 precursor charge in all spectra” [PSI:MS], identificationLevel = "all"; Higher charges may indicate inefficient ionization and/or e.g. preference for longer peptides.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the median of the precursor charge values is calculated and returned.
  • medianCharge, median MS2 precursor charge in identified spectra (MS:4000176), “Median MS2 precursor charge in identified spectra. The used type of identification should be noted in the metadata or analysis methods section of the recording file for the respective run. In case of multiple acceptance criteria (FDR) available in proteomics, PSM-level FDR should be used for better comparability.” [PSI:MS], identificationLevel = "identified"; Higher charges may indicate inefficient ionization and/or e.g. preference for longer peptides in relation to identifiability.

    The metric is calculated as follows:

    1. the Spectra object is filtered according to the MS level,
    2. the precursor charge is obtained,
    3. the median of the precursor charge values is calculated and returned.

4.2 Obtain the metrics in data.frame-format

The most important function to assess the data quality and to calculate the metrics is the calculateMetrics function. The function takes a Spectra or MsExperiment object as input, a character vector of metrics to be calculated, a Boolean value to the filterEmptySpectra argument, and, optionally a list of parameters passed to the quality metrics functions.

The filterEmptySpectra argument specifies if zero-intensity, Inf-intensity or zero-length entries should be removed (filterEmptySpectra = TRUE). By default, the entries are taken as they are (filterEmptySpectra = FALSE). The argument can be set to TRUE to compute metrics that are close to the implementation of the QuaMeter software. Prior to calculating the metrics, the implementation of QuaMeter skips all spectra with defaultArrayLength=0 (in .mzML files) at any MS level.

When passing a Spectra/MsExperiment object to the function, a data.frame returned by calculateMetrics with the metrics specified by the argument metrics. By default, qualityMetrics(object) is taken to specify the calculation of quality metrics. calculateMetrics also accepts a list of parameters passed to the individual quality metrics functions. For each quality metrics functions, the relevant parameters are selected based on the accepted arguments.

Additional arguments can be given to the quality metrics functions. For example, the function ticQuartileToQuartileLogRatio function has the arguments relativeTo, mode, and msLevel. relativeTo specifies to which quantile the log TIC quantile is relatively related to (either to the 1st quantile or the respective previous one). mode (either "TIC_change" or "TIC") specifies if the quantiles are taken from the changes between TICs of scan events or the TICs directly. One Spectra/MsExperiment object may also contain more than one msLevel, e.g. if it also contains information on MS\(^2\) or MS\(^3\) features. If the user adds the arguments relativeTo = "Q1", mode = "TIC", msLevel = c(1L, 2L)), ticQuartileToQuartileLogRatio is run with the parameter combinations relativeTo = "Q1", mode = "TIC", msLevel = c(1L, 2L).

The results based on these parameter combinations are returned and the used parameters are returned as attributes to the returned vector.

Here, we would like to calculate the metrics of all included quality metrics functions (qualityMetrics(object)) and additionally pass the parameter relativeTo = "Q1" and relativeTo = "previous". For computational reasons, we will restrict the calculation of the metrics to the first sample and to RPLC samples.

## subset the Spectra objects
sps_comb_subset <- sps_comb[grep("Sample.1_", sps_comb$dataOrigin), ]

## for RPLC and HILIC
metrics_sps_Q1 <- calculateMetrics(object = sps_comb_subset,
    metrics = qualityMetrics(sps_comb_subset), filterEmptySpectra = FALSE,
    relativeTo = "Q1", msLevel = 1L)
metrics_sps_Q1
##                chromatographyDuration ticQuartersRtFraction.0%
## Sample.1_RPLC                  18.214                        0
## Sample.1_HILIC                 16.000                        0
##                ticQuartersRtFraction.25% ticQuartersRtFraction.50%
## Sample.1_RPLC                 0.08098166                0.08098166
## Sample.1_HILIC                0.34375000                0.34375000
##                ticQuartersRtFraction.75% ticQuartersRtFraction.100%
## Sample.1_RPLC                  0.1495004                          1
## Sample.1_HILIC                 0.5375000                          1
##                rtOverMsQuarters.Quarter1 rtOverMsQuarters.Quarter2
## Sample.1_RPLC                 0.02893379                0.08806413
## Sample.1_HILIC                0.15625000                0.47500000
##                rtOverMsQuarters.Quarter3 rtOverMsQuarters.Quarter4
## Sample.1_RPLC                  0.3102009                         1
## Sample.1_HILIC                 0.6216875                         1
##                ticQuartileToQuartileLogRatio.Q2/Q1
## Sample.1_RPLC                                  NaN
## Sample.1_HILIC                                -Inf
##                ticQuartileToQuartileLogRatio.Q3/Q1
## Sample.1_RPLC                                  NaN
## Sample.1_HILIC                                 NaN
##                ticQuartileToQuartileLogRatio.Q4/Q1 numberSpectra
## Sample.1_RPLC                                  NaN           190
## Sample.1_HILIC                                 NaN           165
##                numberEmptyScans medianPrecursorMz   rtIqr rtIqrRate
## Sample.1_RPLC                 0            198.05 5.10125  18.42686
## Sample.1_HILIC                0            179.10 7.44700  11.14543
##                areaUnderTic areaUnderTicRtQuantiles.25%
## Sample.1_RPLC     655624525                  25612593.2
## Sample.1_HILIC    149945016                    927493.9
##                areaUnderTicRtQuantiles.50% areaUnderTicRtQuantiles.75%
## Sample.1_RPLC                    389734297                   233673968
## Sample.1_HILIC                   100961764                    40996727
##                areaUnderTicRtQuantiles.100% extentIdentifiedPrecursorIntensity
## Sample.1_RPLC                       5692460                          36467.884
## Sample.1_HILIC                      5460065                           8713.906
##                medianTicRtIqr medianTicOfRtRange mzAcquisitionRange.min
## Sample.1_RPLC        7496.006          13858.935                   60.1
## Sample.1_HILIC       2195.400           2086.417                   73.0
##                mzAcquisitionRange.max rtAcquisitionRange.min
## Sample.1_RPLC                  1377.6                  0.986
## Sample.1_HILIC                  784.1                  1.100
##                rtAcquisitionRange.max precursorIntensityRange.min
## Sample.1_RPLC                    19.2                    100.6492
## Sample.1_HILIC                   17.1                    100.0895
##                precursorIntensityRange.max precursorIntensityQuartiles.Q1
## Sample.1_RPLC                    349282909                      1667.3911
## Sample.1_HILIC                    92021751                       300.5038
##                precursorIntensityQuartiles.Q2 precursorIntensityQuartiles.Q3
## Sample.1_RPLC                        6746.195                       75003.90
## Sample.1_HILIC                       2304.383                       33382.25
##                precursorIntensityMean precursorIntensitySd msSignal10xChange
## Sample.1_RPLC               3450655.4             26563228                56
## Sample.1_HILIC               908757.7              7729451                47
##                ratioCharge1over2 ratioCharge3over2 ratioCharge4over2 meanCharge
## Sample.1_RPLC                NaN               NaN               NaN        NaN
## Sample.1_HILIC               NaN               NaN               NaN        NaN
##                medianCharge
## Sample.1_RPLC            NA
## Sample.1_HILIC           NA
## attr(,"chromatographyDuration")
## [1] "MS:4000053"
## attr(,"names1")
## [1] "Q1"
## attr(,"names2")
## [1] "Q2"
## attr(,"names3")
## [1] "Q3"
## attr(,"names4")
## [1] "100%"
## attr(,"names5")
## [1] "100%"
## attr(,"ticQuartersRtFraction")
## [1] "MS:4000054"
## attr(,"rtOverMsQuarters")
## [1] "MS:4000055"
## attr(,"numberSpectra")
## [1] "MS:4000059"
## attr(,"numberEmptyScans")
## [1] "MS:4000099"
## attr(,"areaUnderTic")
## [1] "MS:4000155"
## attr(,"areaUnderTicRtQuantiles")
## [1] "MS:4000156"
## attr(,"mzAcquisitionRange")
## [1] "MS:4000069"
## attr(,"rtAcquisitionRange")
## [1] "MS:4000070"
## attr(,"precursorIntensityRange")
## [1] "MS:4000160"
## attr(,"precursorIntensityQuartiles")
## [1] "MS:4000116"
## attr(,"precursorIntensityMean")
## [1] "MS:4000117"
## attr(,"precursorIntensitySd")
## [1] "MS:4000118"
## attr(,"msSignal10xChange")
## [1] "MS:4000097"
## attr(,"ratioCharge1over2")
## [1] "MS:4000167"
## attr(,"ratioCharge3over2")
## [1] "MS:4000169"
## attr(,"ratioCharge4over2")
## [1] "MS:4000171"
## attr(,"meanCharge")
## [1] "MS:4000173"
## attr(,"medianCharge")
## [1] "MS:4000175"
## attr(,"relativeTo")
## [1] "Q1"
## attr(,"msLevel")
## [1] 1
metrics_sps_previous <- calculateMetrics(object = sps_comb_subset,
    metrics = qualityMetrics(sps_comb_subset), filterEmptySpectra = FALSE,
    relativeTo = "previous", msLevel = 1L)
metrics_sps_previous
##                chromatographyDuration ticQuartersRtFraction.0%
## Sample.1_RPLC                  18.214                        0
## Sample.1_HILIC                 16.000                        0
##                ticQuartersRtFraction.25% ticQuartersRtFraction.50%
## Sample.1_RPLC                 0.08098166                0.08098166
## Sample.1_HILIC                0.34375000                0.34375000
##                ticQuartersRtFraction.75% ticQuartersRtFraction.100%
## Sample.1_RPLC                  0.1495004                          1
## Sample.1_HILIC                 0.5375000                          1
##                rtOverMsQuarters.Quarter1 rtOverMsQuarters.Quarter2
## Sample.1_RPLC                 0.02893379                0.08806413
## Sample.1_HILIC                0.15625000                0.47500000
##                rtOverMsQuarters.Quarter3 rtOverMsQuarters.Quarter4
## Sample.1_RPLC                  0.3102009                         1
## Sample.1_HILIC                 0.6216875                         1
##                ticQuartileToQuartileLogRatio.Q2/Q1
## Sample.1_RPLC                                  NaN
## Sample.1_HILIC                                -Inf
##                ticQuartileToQuartileLogRatio.Q3/Q2
## Sample.1_RPLC                             5.831233
## Sample.1_HILIC                                 Inf
##                ticQuartileToQuartileLogRatio.Q4/Q3 numberSpectra
## Sample.1_RPLC                             8.915513           190
## Sample.1_HILIC                            8.982638           165
##                numberEmptyScans medianPrecursorMz   rtIqr rtIqrRate
## Sample.1_RPLC                 0            198.05 5.10125  18.42686
## Sample.1_HILIC                0            179.10 7.44700  11.14543
##                areaUnderTic areaUnderTicRtQuantiles.25%
## Sample.1_RPLC     655624525                  25612593.2
## Sample.1_HILIC    149945016                    927493.9
##                areaUnderTicRtQuantiles.50% areaUnderTicRtQuantiles.75%
## Sample.1_RPLC                    389734297                   233673968
## Sample.1_HILIC                   100961764                    40996727
##                areaUnderTicRtQuantiles.100% extentIdentifiedPrecursorIntensity
## Sample.1_RPLC                       5692460                          36467.884
## Sample.1_HILIC                      5460065                           8713.906
##                medianTicRtIqr medianTicOfRtRange mzAcquisitionRange.min
## Sample.1_RPLC        7496.006          13858.935                   60.1
## Sample.1_HILIC       2195.400           2086.417                   73.0
##                mzAcquisitionRange.max rtAcquisitionRange.min
## Sample.1_RPLC                  1377.6                  0.986
## Sample.1_HILIC                  784.1                  1.100
##                rtAcquisitionRange.max precursorIntensityRange.min
## Sample.1_RPLC                    19.2                    100.6492
## Sample.1_HILIC                   17.1                    100.0895
##                precursorIntensityRange.max precursorIntensityQuartiles.Q1
## Sample.1_RPLC                    349282909                      1667.3911
## Sample.1_HILIC                    92021751                       300.5038
##                precursorIntensityQuartiles.Q2 precursorIntensityQuartiles.Q3
## Sample.1_RPLC                        6746.195                       75003.90
## Sample.1_HILIC                       2304.383                       33382.25
##                precursorIntensityMean precursorIntensitySd msSignal10xChange
## Sample.1_RPLC               3450655.4             26563228                56
## Sample.1_HILIC               908757.7              7729451                47
##                ratioCharge1over2 ratioCharge3over2 ratioCharge4over2 meanCharge
## Sample.1_RPLC                NaN               NaN               NaN        NaN
## Sample.1_HILIC               NaN               NaN               NaN        NaN
##                medianCharge
## Sample.1_RPLC            NA
## Sample.1_HILIC           NA
## attr(,"chromatographyDuration")
## [1] "MS:4000053"
## attr(,"names1")
## [1] "Q1"
## attr(,"names2")
## [1] "Q2"
## attr(,"names3")
## [1] "Q3"
## attr(,"names4")
## [1] "100%"
## attr(,"names5")
## [1] "100%"
## attr(,"ticQuartersRtFraction")
## [1] "MS:4000054"
## attr(,"rtOverMsQuarters")
## [1] "MS:4000055"
## attr(,"ticQuartileToQuartileLogRatio")
## [1] "MS:4000057"
## attr(,"numberSpectra")
## [1] "MS:4000059"
## attr(,"numberEmptyScans")
## [1] "MS:4000099"
## attr(,"areaUnderTic")
## [1] "MS:4000155"
## attr(,"areaUnderTicRtQuantiles")
## [1] "MS:4000156"
## attr(,"mzAcquisitionRange")
## [1] "MS:4000069"
## attr(,"rtAcquisitionRange")
## [1] "MS:4000070"
## attr(,"precursorIntensityRange")
## [1] "MS:4000160"
## attr(,"precursorIntensityQuartiles")
## [1] "MS:4000116"
## attr(,"precursorIntensityMean")
## [1] "MS:4000117"
## attr(,"precursorIntensitySd")
## [1] "MS:4000118"
## attr(,"msSignal10xChange")
## [1] "MS:4000097"
## attr(,"ratioCharge1over2")
## [1] "MS:4000167"
## attr(,"ratioCharge3over2")
## [1] "MS:4000169"
## attr(,"ratioCharge4over2")
## [1] "MS:4000171"
## attr(,"meanCharge")
## [1] "MS:4000173"
## attr(,"medianCharge")
## [1] "MS:4000175"
## attr(,"relativeTo")
## [1] "previous"
## attr(,"msLevel")
## [1] 1

Alternatively, an MsExperiment object might be passed to calculateMetrics. The function will iterate over the samples (referring to rows in sampleData(msexp))) and calculate the quality metrics on the corresponding Spectras.

4.3 Obtain the metrics in mzQC-format

By default, a data.frame object containing the metric values as entries are returned by the the function calculateMetrics. Alternatively, the function also allows the user to export the metrics in a format defined by the rmzqc package by setting the argument format to "mzQC" (default: format = "data.frame"). In that case, only the metrics that comply to the mzQC specification will be written to the returned object. The object can be exported and validated using the functionality of the rmzqc package (see the documentation of rmzqc for further information).

4.4 Remove empty spectra prior to the calculation

There are in total 541 samples respectively in the objects msexp_rplc and msexp_hilic. To improve the visualization and interpretability, we will only calculate the metrics from the first 20 of these samples.

In this example here, we will remove zero-length and zero-intensity entries prior to calculating the metrics. To do this, we set the filterEmptySpectra argument to TRUE within the calculateMetrics function.

## subset the MsExperiment objects
msexp_rplc_subset <- msexp_rplc[1:20]
msexp_hilic_subset <- msexp_hilic[1:20]

## define metrics
metrics_sps <- c("chromatographyDuration", "ticQuartersRtFraction", "rtOverMsQuarters",
    "ticQuartileToQuartileLogRatio", "numberSpectra", "medianPrecursorMz",
    "rtIqr", "rtIqrRate", "areaUnderTic")

## for RPLC-derived MsExperiment
metrics_rplc_msexp <- calculateMetrics(object = msexp_rplc_subset,
    metrics = qualityMetrics(msexp_rplc_subset), filterEmptySpectra = TRUE,
    relativeTo = "Q1", msLevel = 1L)

## for HILIC-derived MsExperiment
metrics_hilic_msexp <- calculateMetrics(object = msexp_hilic_subset,
    metrics = qualityMetrics(msexp_hilic_subset), filterEmptySpectra = TRUE,
    relativeTo = "Q1", msLevel = 1L)

When passing an MsExperiment object to calculateMetrics a data.frame object is returned with the samples (derived from the rownames of sampleData(msexp)) in the rows and the metrics in columns.

We will show here the objects metrics_rplc_msexp and metrics_hilic_msexp

## [1] "metrics_rplc_msexp"
## [1] "metrics_hilic_msexp"

5 Visualizing the results

The quality metrics can be most easily compared when graphically visualized.

The MsQuality package offers the possibility to graphically display the metrics using the plotMetric and shinyMsQuality functions. The plotMetric function will create one plot based on a single metric. shinyMsQuality, on the other hand, opens a shiny application that allows to browse through all the metrics stored in the object.

As a way of example, we will plot here the number of features. A high number of missing features might indicate low data quality, however, also different sample types might exhibit contrasting number of detected features. As a general rule, only samples of the same type should be compared to adjust for sample type-specific effects.

metrics_msexp <- rbind(metrics_rplc_msexp, metrics_hilic_msexp)
plotMetric(qc = metrics_msexp, metric = "numberSpectra")

Similarly, we are able to display the area under the TIC for the retention time quantiles. This plot gives information on the perceived signal (TIC) for the differnt retention time quantiles and could indicate drifts or interruptions of sensitivity during the run.

plotMetric(qc = metrics_msexp, metric = "ticQuartileToQuartileLogRatio")

Alternatively, to browse through all metrics that were calculated in an interactive way, we can use the shinyMsQuality function.

shinyMsQuality(qc = metrics_msexp)

Appendix

Session information

All software and respective versions to build this vignette are listed here:

## R Under development (unstable) (2024-10-21 r87258)
## Platform: x86_64-pc-linux-gnu
## Running under: Ubuntu 24.04.1 LTS
## 
## Matrix products: default
## BLAS:   /home/biocbuild/bbs-3.21-bioc/R/lib/libRblas.so 
## LAPACK: /usr/lib/x86_64-linux-gnu/lapack/liblapack.so.3.12.0
## 
## locale:
##  [1] LC_CTYPE=en_US.UTF-8       LC_NUMERIC=C              
##  [3] LC_TIME=en_GB              LC_COLLATE=C              
##  [5] LC_MONETARY=en_US.UTF-8    LC_MESSAGES=en_US.UTF-8   
##  [7] LC_PAPER=en_US.UTF-8       LC_NAME=C                 
##  [9] LC_ADDRESS=C               LC_TELEPHONE=C            
## [11] LC_MEASUREMENT=en_US.UTF-8 LC_IDENTIFICATION=C       
## 
## time zone: America/New_York
## tzcode source: system (glibc)
## 
## attached base packages:
## [1] stats4    stats     graphics  grDevices utils     datasets  methods  
## [8] base     
## 
## other attached packages:
## [1] MsExperiment_1.9.0  ProtGenerics_1.39.0 Spectra_1.17.0     
## [4] BiocParallel_1.41.0 S4Vectors_0.45.0    BiocGenerics_0.53.0
## [7] MsQuality_1.7.0     knitr_1.48          BiocStyle_2.35.0   
## 
## loaded via a namespace (and not attached):
##   [1] DBI_1.2.3                   testthat_3.2.1.1           
##   [3] rlang_1.1.4                 magrittr_2.0.3             
##   [5] shinydashboard_0.7.2        clue_0.3-65                
##   [7] matrixStats_1.4.1           compiler_4.5.0             
##   [9] vctrs_0.6.5                 reshape2_1.4.4             
##  [11] stringr_1.5.1               pkgconfig_2.0.3            
##  [13] MetaboCoreUtils_1.15.0      crayon_1.5.3               
##  [15] fastmap_1.2.0               XVector_0.47.0             
##  [17] labeling_0.4.3              utf8_1.2.4                 
##  [19] promises_1.3.0              rmarkdown_2.28             
##  [21] UCSC.utils_1.3.0            purrr_1.0.2                
##  [23] xfun_0.48                   MultiAssayExperiment_1.33.0
##  [25] zlibbioc_1.53.0             cachem_1.1.0               
##  [27] GenomeInfoDb_1.43.0         jsonlite_1.8.9             
##  [29] later_1.3.2                 DelayedArray_0.33.0        
##  [31] parallel_4.5.0              cluster_2.1.6              
##  [33] R6_2.5.1                    RColorBrewer_1.1-3         
##  [35] bslib_0.8.0                 stringi_1.8.4              
##  [37] brio_1.1.5                  GenomicRanges_1.59.0       
##  [39] jquerylib_0.1.4             Rcpp_1.0.13                
##  [41] bookdown_0.41               SummarizedExperiment_1.37.0
##  [43] IRanges_2.41.0              httpuv_1.6.15              
##  [45] Matrix_1.7-1                igraph_2.1.1               
##  [47] tidyselect_1.2.1            abind_1.4-8                
##  [49] yaml_2.3.10                 codetools_0.2-20           
##  [51] curl_5.2.3                  lattice_0.22-6             
##  [53] tibble_3.2.1                plyr_1.8.9                 
##  [55] withr_3.0.2                 Biobase_2.67.0             
##  [57] shiny_1.9.1                 evaluate_1.0.1             
##  [59] ontologyIndex_2.12          pillar_1.9.0               
##  [61] BiocManager_1.30.25         MatrixGenerics_1.19.0      
##  [63] plotly_4.10.4               generics_0.1.3             
##  [65] ggplot2_3.5.1               munsell_0.5.1              
##  [67] scales_1.3.0                R6P_0.3.0                  
##  [69] xtable_1.8-4                glue_1.8.0                 
##  [71] lazyeval_0.2.2              tools_4.5.0                
##  [73] data.table_1.16.2           QFeatures_1.17.0           
##  [75] fs_1.6.4                    grid_4.5.0                 
##  [77] jsonvalidate_1.3.2          tidyr_1.3.1                
##  [79] crosstalk_1.2.1             MsCoreUtils_1.19.0         
##  [81] msdata_0.45.0               colorspace_2.1-1           
##  [83] GenomeInfoDbData_1.2.13     cli_3.6.3                  
##  [85] fansi_1.0.6                 S4Arrays_1.7.0             
##  [87] viridisLite_0.4.2           dplyr_1.1.4                
##  [89] AnnotationFilter_1.31.0     gtable_0.3.6               
##  [91] sass_0.4.9                  digest_0.6.37              
##  [93] SparseArray_1.7.0           htmlwidgets_1.6.4          
##  [95] htmltools_0.5.8.1           lifecycle_1.0.4            
##  [97] httr_1.4.7                  mime_0.12                  
##  [99] rmzqc_0.5.4                 MASS_7.3-61

References

Lee, H.-J., D. M. Kremer, P. Sajjakulnukit, L. Zhang, and C. A. Lyssiotis. 2019. “A Large-Scale Analysis of Targeted Metabolomics Data from Heterogeneous Biological Samples Provides Insights into Metabolite Dynamics.” Metabolomics, 103. https://doi.org/10.1007/s11306-019-1564-8.